NEXIV VMZ-S Series
基本介紹: High Accuracy / High Speed / High Usability 24小時谘詢熱線:High Accuracy / High Speed / High Usability
Magnification, AF and illumination conditions can be finely adjusted for each sample. Suitable for small components used for products such as mechanical, electric/electronic, automotive, and medical devices.
Accuracy within FOV (PF2D,PFV2D) is specified. Nikon designed optics deliver high-quality images and realize accurate measurement data.
Probing error* PF2D 0.8 μm
*Type2 zooming head. Determined by Nikon in-house measurement method. Measurement of the perimeter of circular reference device. It involves stage movement with caliper within the FOV is evenly placed at 25 points on the perimeter.
Probing error of the imaging probe* PFV2D 0.3 μm
*Type2 zooming head. Determined by Nikon in-house measurement method. Measurement of 25 points that are evenly placed on the perimeter of circular device at one locations on the screen (without stage movement).
Throughput has improved by realizing faster short-distance stage movements. This contributes to the shortening of the measurement time.
Along with the exterior design change, the joystick used for stage and optical head movements has been redesigned to make the operation easier and less tiring during long use.
Type 1, 2 and 3, have Nikon 15× zoom optics with a long working distance, wide FOV and a high NA 0.35. They come with LED episcopic and diascopic illuminators, as well as an LED 8-segment ring illuminator with three incident angles.
Type 4 and TZ have Nikon 15× and 120× zoom optics with higher magnification and a higher NA 0.46. Type 4 comes with LED episcopic and diascopic illuminators, as well as an LED 8-segment ring illuminator with a 50-degree incident angle. TZ has LED episcopic, diascopic (for higher magnification) and darkfield illuminators.
Type A features Nikon 10× zoom optics with the widest FOV, the longest working distance and the highest NA 0.11. It comes with LED episcopic and diascopic illuminators, as well as an LED 8-segment ring illuminator with an 18-degree incident angle.
Type 1 | |||||
---|---|---|---|---|---|
Optical magnification | 0.5× | 1× | 2× | 4× | 7.5× |
Total magnification | 18× | 36× | 72× | 144× | 270× |
Field of View (mm) | 9.33 × 7.01 | 4.7 × 3.5 | 2.33 × 1.75 | 1.165 × 0.875 | 0.622 × 0.467 |
Type 2 | |||||
---|---|---|---|---|---|
Optical magnification | 1× | 2× | 4× | 8× | 15× |
Total magnification | 36× | 72× | 144× | 288× | 540× |
Field of View (mm) | 4.7 × 3.5 | 2.33 × 1.75 | 1.165 × 0.875 | 0.582 × 0.437 | 0.311 × 0.233 |
Type 3 | |||||
---|---|---|---|---|---|
Optical magnification | 2× | 4× | 8× | 16× | 30× |
Total magnification | 72× | 144× | 288× | 576× | 1080× |
Field of View (mm) | 2.33 × 1.75 | 1.165 × 0.875 | 0.582 × 0.437 | 0.291 × 0.218 | 0.155 × 0.117 |
Type 4 | |||||
---|---|---|---|---|---|
Optical magnification | 4× | 8× | 16× | 32× | 60× |
Total magnification | 144× | 288× | 576× | 1152× | 2160× |
Field of View (mm) | 1.165 × 0.875 | 0.582 × 0.437 | 0.291 × 0.218 | 0.146 × 0.109 | 0.07 × 0.068 |
Type TZ | ||||||||
---|---|---|---|---|---|---|---|---|
Optical magnification | 1× | 2× | 4× | 7.5× | 16× | 32× | 64× | 120× |
Total magnification | 36× | 72× | 144× | 270× | 576× | 1152× | 2304× | 4320× |
Field of View (mm) | 4.7 × 3.5 | 2.33 × 1.75 | 1.165 × 0.875 | 0.622 × 0.467 | 0.291 × 0.218 | 0.146 × 0.109 | 0.073 × 0.055 | 0.039 × 0.029 |
Type A | |||||
---|---|---|---|---|---|
Optical magnification | 0.35× | 0.6× | 1× | 1.8× | 3.5× |
Total magnification | 12.6× | 21.6× | 36× | 64.8× | 126× |
Field of View (mm) | 13.3 × 10 | 7.8 × 5.8 | 4.7 × 3.5 | 2.6 × 1.9 | 1.33 × 1 |
*Total magnification is that of a video window with 640 × 480 pixels on a 24-inch WUXGA monitor (1920 × 1200 pixels) recommended for the VMZ-S series.
Designed to detect the surfaces of thin transparent materials, the TTL Laser AF offers high speed and high accuracy.
High Accuracy / High Speed / High Usability
Magnification, AF and illumination conditions can be finely adjusted for each sample. Suitable for small components used for products such as mechanical, electric/electronic, automotive, and medical devices.
Accuracy within FOV (PF2D,PFV2D) is specified. Nikon designed optics deliver high-quality images and realize accurate measurement data.
Probing error* PF2D 0.8 μm
*Type2 zooming head. Determined by Nikon in-house measurement method. Measurement of the perimeter of circular reference device. It involves stage movement with caliper within the FOV is evenly placed at 25 points on the perimeter.
Probing error of the imaging probe* PFV2D 0.3 μm
*Type2 zooming head. Determined by Nikon in-house measurement method. Measurement of 25 points that are evenly placed on the perimeter of circular device at one locations on the screen (without stage movement).
Throughput has improved by realizing faster short-distance stage movements. This contributes to the shortening of the measurement time.
Along with the exterior design change, the joystick used for stage and optical head movements has been redesigned to make the operation easier and less tiring during long use.
Type 1, 2 and 3, have Nikon 15× zoom optics with a long working distance, wide FOV and a high NA 0.35. They come with LED episcopic and diascopic illuminators, as well as an LED 8-segment ring illuminator with three incident angles.
Type 4 and TZ have Nikon 15× and 120× zoom optics with higher magnification and a higher NA 0.46. Type 4 comes with LED episcopic and diascopic illuminators, as well as an LED 8-segment ring illuminator with a 50-degree incident angle. TZ has LED episcopic, diascopic (for higher magnification) and darkfield illuminators.
Type A features Nikon 10× zoom optics with the widest FOV, the longest working distance and the highest NA 0.11. It comes with LED episcopic and diascopic illuminators, as well as an LED 8-segment ring illuminator with an 18-degree incident angle.
Type 1 | |||||
---|---|---|---|---|---|
Optical magnification | 0.5× | 1× | 2× | 4× | 7.5× |
Total magnification | 18× | 36× | 72× | 144× | 270× |
Field of View (mm) | 9.33 × 7.01 | 4.7 × 3.5 | 2.33 × 1.75 | 1.165 × 0.875 | 0.622 × 0.467 |
Type 2 | |||||
---|---|---|---|---|---|
Optical magnification | 1× | 2× | 4× | 8× | 15× |
Total magnification | 36× | 72× | 144× | 288× | 540× |
Field of View (mm) | 4.7 × 3.5 | 2.33 × 1.75 | 1.165 × 0.875 | 0.582 × 0.437 | 0.311 × 0.233 |
Type 3 | |||||
---|---|---|---|---|---|
Optical magnification | 2× | 4× | 8× | 16× | 30× |
Total magnification | 72× | 144× | 288× | 576× | 1080× |
Field of View (mm) | 2.33 × 1.75 | 1.165 × 0.875 | 0.582 × 0.437 | 0.291 × 0.218 | 0.155 × 0.117 |
Type 4 | |||||
---|---|---|---|---|---|
Optical magnification | 4× | 8× | 16× | 32× | 60× |
Total magnification | 144× | 288× | 576× | 1152× | 2160× |
Field of View (mm) | 1.165 × 0.875 | 0.582 × 0.437 | 0.291 × 0.218 | 0.146 × 0.109 | 0.07 × 0.068 |
Type TZ | ||||||||
---|---|---|---|---|---|---|---|---|
Optical magnification | 1× | 2× | 4× | 7.5× | 16× | 32× | 64× | 120× |
Total magnification | 36× | 72× | 144× | 270× | 576× | 1152× | 2304× | 4320× |
Field of View (mm) | 4.7 × 3.5 | 2.33 × 1.75 | 1.165 × 0.875 | 0.622 × 0.467 | 0.291 × 0.218 | 0.146 × 0.109 | 0.073 × 0.055 | 0.039 × 0.029 |
Type A | |||||
---|---|---|---|---|---|
Optical magnification | 0.35× | 0.6× | 1× | 1.8× | 3.5× |
Total magnification | 12.6× | 21.6× | 36× | 64.8× | 126× |
Field of View (mm) | 13.3 × 10 | 7.8 × 5.8 | 4.7 × 3.5 | 2.6 × 1.9 | 1.33 × 1 |
*Total magnification is that of a video window with 640 × 480 pixels on a 24-inch WUXGA monitor (1920 × 1200 pixels) recommended for the VMZ-S series.
High Accuracy / High Speed / High Usability
Magnification, AF and illumination conditions can be finely adjusted for each sample. Suitable for small components used for products such as mechanical, electric/electronic, automotive, and medical devices.
Accuracy within FOV (PF2D,PFV2D) is specified. Nikon designed optics deliver high-quality images and realize accurate measurement data.
Probing error* PF2D 0.8 μm
*Type2 zooming head. Determined by Nikon in-house measurement method. Measurement of the perimeter of circular reference device. It involves stage movement with caliper within the FOV is evenly placed at 25 points on the perimeter.
Probing error of the imaging probe* PFV2D 0.3 μm
*Type2 zooming head. Determined by Nikon in-house measurement method. Measurement of 25 points that are evenly placed on the perimeter of circular device at one locations on the screen (without stage movement).
Throughput has improved by realizing faster short-distance stage movements. This contributes to the shortening of the measurement time.
Along with the exterior design change, the joystick used for stage and optical head movements has been redesigned to make the operation easier and less tiring during long use.
Type 1, 2 and 3, have Nikon 15× zoom optics with a long working distance, wide FOV and a high NA 0.35. They come with LED episcopic and diascopic illuminators, as well as an LED 8-segment ring illuminator with three incident angles.
Type 4 and TZ have Nikon 15× and 120× zoom optics with higher magnification and a higher NA 0.46. Type 4 comes with LED episcopic and diascopic illuminators, as well as an LED 8-segment ring illuminator with a 50-degree incident angle. TZ has LED episcopic, diascopic (for higher magnification) and darkfield illuminators.
Type A features Nikon 10× zoom optics with the widest FOV, the longest working distance and the highest NA 0.11. It comes with LED episcopic and diascopic illuminators, as well as an LED 8-segment ring illuminator with an 18-degree incident angle.
Type 1 | |||||
---|---|---|---|---|---|
Optical magnification | 0.5× | 1× | 2× | 4× | 7.5× |
Total magnification | 18× | 36× | 72× | 144× | 270× |
Field of View (mm) | 9.33 × 7.01 | 4.7 × 3.5 | 2.33 × 1.75 | 1.165 × 0.875 | 0.622 × 0.467 |
Type 2 | |||||
---|---|---|---|---|---|
Optical magnification | 1× | 2× | 4× | 8× | 15× |
Total magnification | 36× | 72× | 144× | 288× | 540× |
Field of View (mm) | 4.7 × 3.5 | 2.33 × 1.75 | 1.165 × 0.875 | 0.582 × 0.437 | 0.311 × 0.233 |
Type 3 | |||||
---|---|---|---|---|---|
Optical magnification | 2× | 4× | 8× | 16× | 30× |
Total magnification | 72× | 144× | 288× | 576× | 1080× |
Field of View (mm) | 2.33 × 1.75 | 1.165 × 0.875 | 0.582 × 0.437 | 0.291 × 0.218 | 0.155 × 0.117 |
Type 4 | |||||
---|---|---|---|---|---|
Optical magnification | 4× | 8× | 16× | 32× | 60× |
Total magnification | 144× | 288× | 576× | 1152× | 2160× |
Field of View (mm) | 1.165 × 0.875 | 0.582 × 0.437 | 0.291 × 0.218 | 0.146 × 0.109 | 0.07 × 0.068 |
Type TZ | ||||||||
---|---|---|---|---|---|---|---|---|
Optical magnification | 1× | 2× | 4× | 7.5× | 16× | 32× | 64× | 120× |
Total magnification | 36× | 72× | 144× | 270× | 576× | 1152× | 2304× | 4320× |
Field of View (mm) | 4.7 × 3.5 | 2.33 × 1.75 | 1.165 × 0.875 | 0.622 × 0.467 | 0.291 × 0.218 | 0.146 × 0.109 | 0.073 × 0.055 | 0.039 × 0.029 |
Type A | |||||
---|---|---|---|---|---|
Optical magnification | 0.35× | 0.6× | 1× | 1.8× | 3.5× |
Total magnification | 12.6× | 21.6× | 36× | 64.8× | 126× |
Field of View (mm) | 13.3 × 10 | 7.8 × 5.8 | 4.7 × 3.5 | 2.6 × 1.9 | 1.33 × 1 |
*Total magnification is that of a video window with 640 × 480 pixels on a 24-inch WUXGA monitor (1920 × 1200 pixels) recommended for the VMZ-S series.
Designed to detect the surfaces of thin transparent materials, the TTL Laser AF offers high speed and high accuracy.